Self Tests
Performance Verification
4Ć6
Figure 4Ć2:ăCalibrations Menu
b. Run the adjustments routine: Select Execute from the side menu.
This executes the AWG2005 calibration routines automatically.
c. Wait: The internal calibration does an exhaustive verification of
proper AWG2005 function. While this verification progresses, the
clock icon appears on screen. When finished, the resulting status
will appear on the screen.
d. Confirm that no failures are found: Verify that no failures are found
and reported onĆscreen. If the calibration displays FAIL as the result,
use the instructions in section 6, Troubleshooting, to identify the
faulty module. If an error code is displayed, contact the nearest
Tektronix representative.
2. Return to regular service: Push any button (other than UTILITY) in the
MENU column to exit the calibration menu.