Tektronix AWG2005 Stereo System User Manual


 
Performance Tests
AWG2005 Service Manual
4Ć43
Vertical CH1
Coupling DC
Scale 0.2ĂV/div.
Input Impedance 50Ă
Horizontal
Sweep 500õs/div.
Trigger
Source CH1
Coupling DC
Slope Positive
Level 0ĂmV
Mode Auto
d. Set function generator controls:
Function Square
Mode Continuous
Parameter
Frequency 1ĂMHz
Amplitude 2.0ĂV
Offset 1.0ĂV
Output Off
2. Select the AWG2005 waveform file and set AWG2005 controls:
a. Initialize AWG2005 controls: Push UTILITY!Misc!Config...!ReĆ
set to Factory!O.K.
b. Select waveform file:
H Push SETUP!Waveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
H Turn the general purpose knob to highlight the EXT_CLK.WFM
file.
H Push ENTER to select the file.
3. Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED
above the CH1 output connector is on.
4. Check the external CLOCK IN threshold level:
a. Enable function generator output: Turn on function generator output.
b. Check the level: Check that the waveform displayed on the oscilloĆ
scope has an amplitude of 5 divisions and a stable display of 5
cycles.
5. Turn off equipment output and disconnect test hookup:
a. Disable function generator output: Turn off function generator output.
b. Remove connections: Disconnect all connections to the AWG2005.