Before installing the AWG2005, note these precautions:
WARNING
To avoid equipment failure and potential fire or personal shock
hazards, do not exceed the maximum rated operating voltage of
250ĂV between the voltageĆtoĆground (earth) and either pole of the
power source. The AWG2005 operates from a singleĆphase power
source and has a threeĆwire power cord with a twoĆpole, threeĆterĆ
minal grounding plug. Also, before making connection to the power
source, be sure the AWG2005 has a suitable twoĆpole, threeĆtermiĆ
nal groundingĆtype plug.
To avoid personal shock hazard, do not contact conductive parts.
All accessible conductive parts are directly connected through the
grounding conductor of the power cord to the grounded (earthing)
contact of the power plug. The AWG2005 is safety Class 1 equipĆ
ment (IEC designation).
To avoid personal shock hazard, do not defeat the grounding
connection. Insert the power input plug only in a mating receptacle
with a grounding contact where earth ground has been verified by a
qualified service person. Also, for electricalĆshock protection, make
the grounding connection before making connection to the
AWG2005 input or output terminals.
The AWG2005 ships with the required power cord as ordered by the cusĆ
tomer. Table 2Ć1 gives the colorĆcoding of the conductors in the power cord.
Figure 2Ć1 shows information on the available power cords.
TableĂ2Ć1:ăPowerĆcord Conductor Identification
Conductor
Color Alternate Color
Ungrounded (Line) Brown Black
Grounding (Neutral) Light Blue White
Grounding (Earthing) Green/Yellow Green