ICS-2100 Ion Chromatography System
84 Doc. 065291-01 3/09
This error occurs if you turn on the suppressor and the ICS-2100 cannot
establish a connection with the suppressor.
To troubleshoot:
1. Check the suppressor cable connection (see Section 5.11
).
2. If the error persists, contact Dionex for assistance. There may be a
problem in the suppressor controller electronics. The ICS-2100
electronics components cannot be serviced by the user.
This error may be caused by an expended or dirty suppressor or by a
malfunction in the suppressor controller electronics.
To troubleshoot:
1. Follow the instructions in the suppressor manual to regenerate the
suppressor. Suppressor manuals are included on the Dionex Reference
Library CD-ROM (P/N 053891).
2. Follow the instructions in the suppressor manual to clean the suppressor.
3. If a malfunction in the suppressor controller is suspected, contact Dionex
for assistance. The ICS-2100 electronics components cannot be serviced
by the user.
SUPPRESSOR NOT CONNECTED
SUPPRESSOR OVER CURRENT