Oritron ICS-2100 Home Theater System User Manual


 
ICS-2100 Ion Chromatography System
66 Doc. 065291-01 3/09
4. Click the Autozero button on the Control panel or the touch screen
HOME page.
5. Inject the sample (see Section 3.10.4
).
6. In Chromeleon, monitor the chromatogram on the Control panel and
select Control > Acquisition Off when sample data is collected.
NOTE Data from Chromeleon manual processing is
saved in the MANUAL sequence folder under
the local datasource. This data is overwritten
each time a new manual sample is processed. To
save the data from a manual run, select the
MANUAL folder, select File > Save As, and
enter a new name for the sequence.
7. If you are controlling the ICS-2100 from the touch screens, select the
PLOT button on the HOME page to view a plot of the conductivity
data.
3.11.2 Automatic (Batch) Sample Processing
You can use Chromeleon or Chromeleon Xpress to create a list of samples
(a sequence) to be processed automatically. For each sample, the
sequence includes the following:
A program with commands and parameters for controlling the ICS-
2100 and autosampler (if used), and for acquiring sample data.
A quantification method for peak identification and area
determination.
Additional sample processing parameters (sample name, sample type,
injection volume, etc.).
After creating the sequence, you can start batch processing.