ICS-2100 Ion Chromatography System
160 Doc. 065291-01 3/09
1. Disconnect the TO INJ VALVE IN - P/ELUENT OUT line from the EGC
CO
3
Mixer inlet.
2. Locate the yellow 0.5 mL/min, 7 MPa (1000 psi) backpressure coil
(P/N 053765) and a 10-32 to 10-32 coupler (P/N 042627) in the ICS-
2100 Ship Kit (P/N 064375).
3. Use the coupler to connect the
TO INJ VALVE IN - P/ELUENT OUT line
to one end of the backpressure coil (see Figure 5-56
).
4. Direct the free end of the backpressure coil to a temporary waste
container.
Figure 5-56. Plumbing Connections for Conditioning the
EluGen Cartridge with Carbonate Eluent
EGC Cable
Temporary
Waste
Container
Back-
pressure
Coil
TO INJ VALVE
IN - P/
ELUENT OUT