3 Operation and Maintenance
Doc. 065291-01 3/09 55
The STATUS page (see Figure 3-7) also displays the amount of liquid
remaining in the reservoir and an estimate of how many hours remain
before the eluent reservoir is empty.
In order for the level displayed in the Eluent Bottle box and gauge to
be accurate, you must enter the level when the reservoir is filled. The
ICS-2100 does not automatically detect when the reservoir is filled,
nor when it is empty.
Figure 3-5. Home Page: Setting the Eluent Level
Figure 3-6. Pump Settings Windows: Setting the Eluent Level
Enter the volume of
liquid in the reservoir
Note: The Eluent Level
field is updated as the
liquid is depleted.
Enter the volume
of liquid in the
reservoir (in liters)
Note: The Eluent
Bottle field and slider
gauge are updated as
the liquid is depleted.