4.2 Troubleshooting Error Messages
This error occurs when the auxiliary power supply is configured as active, but
no device is connected.
To troubleshoot:
1. Connect the cable for the auxiliary device to the
EGC-2 power supply
connector.
2. If there is no auxiliary device, open the Chromeleon Server Configuration
program and remove this option from the ICS-2100 timebase.
This error occurs when the device connected to the auxiliary power supply is
attempting to draw more current than is allowed (200 mA) or is 20% higher
than the set current.
To troubleshoot:
1. Check the cable connecting the device to the power supply for damage or
a faulty connection.
2. If the error persists, refer to the manual provided by the device vendor for
assistance.
AUXILIARY POWER SUPPLY DISCONNECTED
AUXILIARY POWER SUPPLY OVER CURRENT