Oritron ICS-2100 Home Theater System User Manual


 
5•Service
Doc. 065291-01 3/09 185
17. Reconnect the TO SRS/AES REGEN OUT line to the suppressor REGEN
OUT
port.
18. Reconnect the mixer outlet line to port
P (2) on the injection valve.
5.23 Replacing the EGC Holder and Degas Assembly
5.23.1 Disconnecting and Removing the EluGen Cartridge
1. Turn off the power to the ICS-2100.
2. Disconnect the following EGC holder liquid lines:
Disconnect the line labeled TO INJ VALVE IN - P/ELUENT OUT
from either port
P (2) on the injection valve, or if an EGC CO
3
Mixer inlet is installed, from the mixer inlet.
If a CR-TC is installed, disconnect the line labeled TO SRS/AES
REGEN OUT
from the REGEN OUT port on the suppressor.
Disconnect the line labeled TO PUMP/DAMPER from the pump
damper.
Disconnect the line labeled WASTE-GAS SEPARATOR from the
gas separator waste tube assembly installed on the waste
container.
3. Disconnect the EluGen cartridge cable, and the CR-TC or EPM cable
(if installed) from their respective connectors on the ICS-2100 top
cover.
The EluGen cartridge contains one of the following: a corrosive base
(KOH, LiOH, or NaOH), or a corrosive acid (MSA). Wear protective
eyewear and gloves when handling the cartridge.
La cartouche d'EluGen contient un de ce qui suit: une base corrosive
(KOH, LiOH, ou NaOH), ou un acide corrosif (MSA). Porter des
lunettes et des gants protectives en manipulant la cartouche.
Die EluGen-Kartusche enthält eine korrosive Base (KOH, LiOH oder
NaOH), oder eine korrosive Säure (MSA). Tragen Sie daher beim
Umgang mit der Kartusche eine Schutzbrille und Handschuhe.