15. Turn the EluGen cartridge over (with the fittings facing down). Shake
the EluGen cartridge vigorously, and tap it with the palm of your
hand 10 to 15 times. Check that all bubbles trapped in the eluent
generation chamber are dislodged.
16. Slide the EluGen cartridge down into the
holder until it stops (see Figure 5-52
).
Make sure the cable remains out of the
holder.
NOTE To avoid crimping the lines, do not
twist the cartridge when you slide
it into the holder. The lines are
designed to coil inside the holder.
Figure 5-52. EluGen Cartridge Installed in the Holder
The EluGen cartridge contains one of the following: a corrosive base
(KOH, LiOH, or NaOH), a corrosive acid (MSA), or a concentrated
K
2
CO
3
solution. Wear protective eyewear and gloves when handling
the cartridge.
La cartouche d'EluGen contient un de ce qui suit: une base corrosive
(KOH, LiOH, ou NaOH), un acide corrosif (MSA), ou une solution
concentrée de K
2
CO
3
. Porter des lunettes et des gants protectives en
manipulant la cartouche.
Die EluGen-Kartusche enthält eine korrosive Base (KOH, LiOH oder
NaOH), oder eine korrosive Säure (MSA). Tragen Sie daher beim
Umgang mit der Kartusche eine Schutzbrille und Handschuhe.