12. Align the new valve on the component panel with the liquid ports facing up
and attach it with the two screws removed in Step 5
. Reconnect the liquid
lines.
13. Turn on the power to the ICS-2100.
14. Prime the pump (see Section 5.17
).
5.15 Replacing the Leak Sensor
1. Turn off the power to the ICS-
2100.
2. Open the ICS-2100 front door.
3. Loosen the screw on the front of
the leak sensor (see
Figure 5-29
). Note: The screw
remains attached to the sensor.
Figure 5-29. Leak Sensor
4. Remove the leak sensor from
the component panel and pull
the cable out of the opening in
the panel.
5. The cable attached to the leak
sensor is connected to a
matching cable inside the ICS-
2100. Continue pulling the cable
until the connectors for the two
cables are outside of the panel
(see Figure 5-30
).
6. Disconnect the two cables.
Figure 5-30. Leak Sensor Cable
Leak
Sensor
Loosen
Screw
Disconnect
Cables