ICS-2100 Ion Chromatography System
80 Doc. 065291-01 3/09
This error indicates a problem in the ICS-2100 electronics.
To troubleshoot:
Contact Dionex for assistance. The ICS-2100 electronics components cannot
be serviced by the user.
The leak sensor is installed in the drip tray at the bottom of the component
panel (see Figure 2-8
). If liquid accumulates in the tray, the sensor signals the
problem and this error message appears.
To troubleshoot:
1. Locate the source of the leak by visually inspecting the tubing, fittings,
and components on the component panel. Refer to Section 4.3
.
2. Tighten fittings or replace tubing and fittings as required. Refer to
Section 4.3
for detailed troubleshooting of various types of leaks.
3. After fixing the leak, dry the drip tray thoroughly to prevent the leak
sensor from triggering additional error messages.
If the injection valve fails to switch position within 1 second of being toggled,
the ICS-2100 Moduleware reports an error to Chromeleon and this error
message appears.
To troubleshoot:
1. If a sequence is being executed, terminate the sequence by selecting
Batch > Stop in Chromeleon.
2. Try to toggle the valve from Load to Inject by clicking the Load
and
Inject buttons on the ICS-2100 Control panel in Chromeleon or on the
touch screen
HOME page.
HARDWARE NOT PRESENT
LEAK SENSOR WET
LOAD/INJECT VALVE ERROR