5•Service
Doc. 065291-01 3/09 193
Connect the line labeled TO INJ VALVE IN - P/ELUENT OUT to the
EGC CO
3
Mixer inlet.
Connect the line labeled TO PUMP/DAMPER to the pulse damper.
4. Direct the
WASTE, GAS SEPARATOR line from the top of the ICS-
2100 to the rear panel. Snap the line onto one of the tubing clips on
the rear panel. Connect the line to the gas separator waste tube (see
Section B.10.1
).
5. Go on to Section 5.23.5
.
5.23.5 Reinstalling the EluGen Cartridge
1. Pull the coiled black EGC IN and EGC OUT lines out from inside the
EGC holder. Disconnect the
EGC IN line from the coupler that
connects it to the
EGC OUT line.
2. Connect the
EGC IN
line to the EluGen
cartridge
INLET port
(see Figure 5-82
).
3. Remove the coupler
and connect the
EGC
OUT
line to the
EluGen cartridge
OUTLET port.
Figure 5-82. EluGen Cartridge Inlet and Outlet Port Connections
4. Turn over the EluGen cartridge (fittings facing down). Check for any
bubbles in the eluent generation chamber, and if necessary, shake and
tap the EluGen cartridge to remove them.
5. Reinstall the EluGen cartridge in the holder.
6. Reconnect the EluGen cartridge cable and the CR-TC or EPM cable
(if installed).
Remove the cap from the luer adapter on the EluGen cartridge vent
opening and re-connect the
EGC VENT line.
EGC OUT
OUTLET
Port
EGC IN
INLET
Port