Oritron ICS-2100 Home Theater System User Manual


 
ICS-2100 Ion Chromatography System
240 Doc. 065291-01 3/09
The outputs can be used to control functions in external devices such as an
autosampler or another Dionex module. When connected to a controlling device,
the inputs can be programmed to perform the following ICS-2100 functions:
Switch the injection valve position (load/inject)
Perform an autozero command (set the conductivity to zero)
Turn the pump on and off (also turns the suppressor on and off)
Automatically turn on the optional auxiliary power supply simultaneously
with the pump flow
Send a chart mark signal to the analog output. The mark is 10% of the full-
scale voltage, and the duration is 0.5 seconds. A mark can be used, for
example, to indicate the injection.
Relay outputs 1 and 2 can be programmed to switch any low-voltage control.
Switched current must be less than 200 mA and 42 V peak.
Connecting a TTL or Relay
1. Locate the twisted pair of
wires (P/N 043598) and the
12-position connector plug
(P/N 923687) (see
Figure C-2
) in the ICS-2100
Ship Kit (P/N 064375).
2. Follow these basic steps to
connect the TTL or relays.
a. For each relay or TTL to
be used, connect an
active wire (red) and a ground wire (black) to the 12-position connector
plug at the appropriate pin locations. Refer to Figure C-1
or the label on
the ICS-2100 rear panel for the connector pin assignments.
To attach a wire to the plug, strip the end of the wire, insert it into the
plug, and use a screwdriver to tighten the locking screw. If necessary,
multiple ground wires can be attached to a single TTL input/output
ground pin.
When attaching wires to the connector plug, be careful not to allow
stray strands of wire to short to the adjoining position on the
connector.
Figure C-2. 12-Position Connector Plug
Position 1
Position 12
Lockin
g
Screws