ICS-2100 Ion Chromatography System
226 Doc. 065291-01 3/09
B.8 Eluent Generator Page
Use the ELUENT GENERATOR page to set the eluent concentration of each
installed EluGen cartridge (EGC). You can also enter the serial number and view
the expiration date and remaining life of each cartridge. This page also provides
on/off control of a Continuously Regenerated Trap Column (CR-TC). To open the
page, touch the
EGC button on the HOME page or select EGC from the menu of
pages.
B.8.1 EGC Serial Number
The Serial Number field displays the serial number of the EluGen
cartridge currently installed. After installing a new EluGen cartridge,
enter the cartridge serial number. If you use Chromeleon to control the
ICS-2100, enter the serial number in the Chromeleon Server
Configuration, not on the
ELUENT GENERATOR page. If you enter the
serial number first in Chromeleon, the number is displayed on the
ELUENT GENERATOR page. However, if you enter the serial number on
the
ELUENT GENERATOR page, it will not be available in Chromeleon.
Figure B-16. Eluent Generator Page