Oritron ICS-2100 Home Theater System User Manual


 
ICS-2100 Ion Chromatography System
20 Doc. 065291-01 3/09
Suppressor
The suppressor reduces the eluent conductivity and enhances the
conductivity of the sample ions, thereby increasing detection sensitivity.
Either an AES
®
Atlas Electrolytic Suppressor, SRS
®
Self-Regenerating
Suppressor, or MMS
MicroMembrane
Suppressor can be used with
the ICS-2100. See Section 2.4.9
for details about the suppressor.
Separator and Guard Columns
Both the separator and guard columns are packed with resin and perform
the separation of the sample ions. The main function of the guard column
is to trap contaminants and remove particulates that might damage the
separator column.
Column Heater
The column heater controls the temperature of the separator and guard
columns. The temperature can be set to between 30 and 60 °C; however,
it must be set to at least 5 °C above the ambient temperature. See
Section 2.4.8
for details about the column heater.
Eluent Valve
The eluent valve controls the flow from the eluent reservoir. The eluent
valve opens automatically when the pump is started and closes when the
pump is turned off.
Screen Adjustment Knob
Use this knob to change the LCD touch screen contrast.
Tubing Chases
The upper tubing chase routes tubing from the top cover to the component
panel. The lower tubing chase routes tubing from the component panel,
through the interior of the ICS-2100, to the rear panel.