Oritron ICS-2100 Home Theater System User Manual


 
5•Service
Doc. 065291-01 3/09 107
3. On the Chromeleon Wellness Panel under Pump Flow Rate
Calibration, select Reset Cal, or touch
Reset Cal on the touch
screen.
4. Select
1 mL/min to set the flow rate and begin pumping deionized
water.
5. Allow the pump to stabilize for 20 minutes.
6. If you are using the Wellness Panel, click
5.00 min to start the timer.
If you are using the touch screen, start an external timer. Immediately
start collecting water into the tared beaker.
7. Collect the water for exactly 5.00 minutes.
8. Enter the weight of the water (in grams).
9. If you are using the Wellness Panel, click
Calibrate to download the
value to the ICS-2100.
10. To recheck the value, wait at least 15 minutes, and then repeat the
calibration.
11. If you are using the Wellness panel, click
Log to record the new
calibration value in the Audit Trail.
12. If you performed the calibration from the ICS-2100 touch screen,
record the new calibration values in the Chromeleon Wellness
database. To do this, open the Wellness Panel (see Section 5.1.1
) and
click Upload.