4. Remove the EluGen
cartridge
VENT line
(see Figure 5-35
)
from the EluGen
cartridge by
unscrewing the luer
lock from the luer
adapter.
If necessary, use a
wrench to hold the
luer adapter in place
while unscrewing
the vent line.
Figure 5-35. EluGen Cartridge Vent Line and Luer Fitting
5. Install the luer cap (P/N 053981), provided with the EluGen cartridge,
onto the luer adapter. This prevents leaks from the vent opening when
you turn over the EluGen cartridge (see Step 7)
.
6. Lift the EluGen cartridge straight up and out of the cartridge holder.
7. Turn the cartridge upside down and set it in the EluGen cartridge
service area on top of the ICS-2100 (see Figure 5-36
and
Figure 5-37
).
Figure 5-36. EluGen Cartridge Service Area
Luer Lock
Luer Adapter
EGC
Vent
Line
EluGen Cartridge
AreaService