ICS-2100 Ion Chromatography System
174 Doc. 065291-01 3/09
14. Reconnect the TO INJ VALVE IN - P/ELUENT OUT line to the EGC CO
3
Mixer inlet.
15. Reconnect the
TO SRS/AES REGEN OUT line to the suppressor REGEN
OUT
port.
5.22 Replacing the EGC CO
3
Mixer
5.22.1 Installing the New EGC CO
3
Mixer
1. Turn off the pump from the ICS-2100 Control panel in Chromeleon
or from the touch screen
HOME page.
2. Disconnect the inlet and outlet lines (labeled
TO EGC CO3 MIXER IN
and
TO EGC CO3 MIXER OUT) from the existing mixer and pull the
mixer off the mounting clips.
3. Orient the new EGC CO
3
Mixer (P/N 061686) with the outlet
pointing upward and push it onto the mounting clips (see
Figure 5-67
). The mixer outlet must point upward to ensure thorough
mixing of the eluent.