4 Troubleshooting
Doc. 065291-01 3/09 91
Flow rate is set to 0
Set the flow rate from the ICS-2100 Control panel in Chromeleon or from the
touch screen
HOME page.
Eluent valve is closed
Open the eluent valve from the ICS-2100 Control panel in Chromeleon or
from the touch screen
PUMP page.
Pump not primed
Prime the pump (see Section 5.17
).
Broken pump piston
Replace the piston (see Section 5.8
).
Eluent generator degas tubing assembly ruptured
Replace the EluGen cartridge holder (see Section 5.19
).
4.7 Erratic Flow/Pressure Reading
Pump needs priming
Prime the pump (see Section 5.17
).
Figure 4-2. Waste and Priming Valves
Waste Valve
Priming Valve