ICS-2100 Ion Chromatography System
192 Doc. 065291-01 3/09
8. Turn the EGC holder right side up and set it in place on top of the
ICS-2100.
9. Slide the EPM mounting bracket and connected EPM onto the side of
the EGC holder.
10. Feed the following liquid lines through the tubing chase to the ICS-
2100 front component panel.
TO SRS/AES REGEN OUT
TO INJ VALVE IN - P/ELUENT OUT
TO PUMP/DAMPER
11. Connect the lines to their respective locations:
Connect the line labeled TO SRS/AES REGEN OUT to the REGEN
OUT
port on the suppressor.
Connect the line labeled TO INJ VALVE IN - P/ELUENT OUT to the
EGC CO
3
Mixer inlet.
Connect the line labeled TO PUMP/DAMPER to the pulse damper.
12. Direct the
WASTE, GAS SEPARATOR line from the top of the ICS-
2100 to the rear panel. Snap the line onto one of the tubing clips on
the rear panel. Connect the line to the gas separator waste tube (see
Section B.10.1
).
13. Go on to Section 5.23.5
.
5.23.4 Installing the New EGC Holder Without a CR-TC or EPM
Follow the instructions in this section if a CR-TC or EPM is not installed.
1. Set the new EGC holder in place on top of the ICS-2100.
2. Feed the following liquid lines through the tubing chase to the ICS-
2100 front component panel.
TO SRS/AES REGEN OUT
TO INJ VALVE IN - P/ELUENT OUT
TO PUMP/DAMPER
3. Connect the lines to their respective locations:
Connect the line labeled TO SRS/AES REGEN OUT to the REGEN
OUT
port on the suppressor.