ICS-2100 Ion Chromatography System
104 Doc. 065291-01 3/09
2. On the Wellness Panel under Electric Conductivity Cell
Calibration, click Offset Cal, or touch
Run Offset Cal on the ICS-
2100 touch screen.
3. When the offset calibration is complete, click Slope Cal on the
Wellness Panel, or touch
Run Slope Cal on the ICS-2100 touch
screen.
4. If you are using the Wellness Panel, click Log to record the new
calibration values in the Audit Trail.
5. Disconnect the pump output line from the injection valve.
6. Disconnect the line from the suppressor
ELUENT OUT port to the cell
inlet and connect the pump output line directly to the cell inlet.
7. Disconnect the line from the suppressor
REGEN IN port to the cell
outlet. Direct the cell outlet line to a waste container.
8. Verify that the backpressure is at least 7 MPa (1000 psi).
9. If you are using the touch screens to perform the calibration, open the
CONDUCTIVITY CELL page (see Figure 5-4).
10. On the Wellness Panel under Conductivity Cell Calibration, click
Cell 35 °C, or touch
35 C on the ICS-2100 touch screen. Allow the
cell to reach this temperature, and then wait an additional 5 minutes to
Figure 5-4. Conductivity Cell Page