4.9 Peak “Ghosting”
“Ghosting” is the appearance of extraneous peaks in a chromatogram. These may
be late-eluting peaks from a previous injection. They may also be the result of a
contaminated standard or eluent, or a malfunctioning injection valve. These peaks
may co-elute with peaks of interest, resulting in nonreproducible peak
heights/areas.
Insufficient time between sample injections
Wait until the previous sample has been completely eluted before making
another injection.
Insufficient flush between samples
Flush the sample loop with at least 10 loop volumes of deionized water or
sample between sample injections.
Incorrect or contaminated standards
Remake standards.
Incorrect or contaminated eluent
Remake the eluent. Refer to the instructions in the column manual. The
manual is included on the Dionex Reference Library CD-ROM (P/N 053891).
Verify that the correct eluent concentration is selected.
Install or replace the end-line filter on the end of the deionized water line (see
Section 3.4.1
).
Malfunctioning injection valve or auxiliary valve
1. Rebuild the valve (see Section 5.4
).
2. If the valve leak persists, contact Dionex for assistance.