Oritron ICS-2100 Home Theater System User Manual


 
ICS-2100 Ion Chromatography System
170 Doc. 065291-01 3/09
5.21.1 Recording the EPM Serial Number in Chromeleon
The serial number is printed on a label on the EPM.
1. Open the Server Configuration program (select Start > All
Programs > Chromeleon > Server Configuration).
2. Double-click the ICS-2100 system icon. The properties dialog box for
the ICS-2100 appears.
3. Click the Eluent Generator tab (see Figure 5-44
).
4. Select the EGC-2 check box.
5. Enter the serial number of the EPM.
6. Verify that the cartridge type is
pH Mod. If not, make sure the serial
number was entered correctly.
7. Click OK.
Figure 5-64. Eluent Generator Configuration Properties Example