Oritron ICS-2100 Home Theater System User Manual


 
ICS-2100 Ion Chromatography System
58 Doc. 065291-01 3/09
reminder message appears. The pump will begin pumping at approximately
3mL/min.
4. Continue priming the ICS-2100 until all air and previous eluent are purged
and no air bubbles are exiting the waste line.
5. Press
Pump Off.
6. Close the waste valve. Do not overtighten. The pump is now ready for
operation.
3.7 Set System Operating Conditions
NOTE This section is an overview of the steps needed to start
up the system and begin running samples. Actual
operating parameters (flow rate, cell heater
temperature, suppressor current, etc.) depend on the
application to be run. Refer to the column manual for
the required parameters for your application.
Set or verify system operating parameters from either the Chromeleon Control
panel or from the touch screen
HOME page. Operating parameters can also be set
automatically by loading a Chromeleon sequence.
NOTE Clicking the System Startup button on the Chromeleon
Control panel starts the pump and suppressor. The flow
rate and suppressor current settings that were in effect
when the system was shut down are restored.
NOTE Clicking the System Startup button on the Chromeleon
Control panel starts the pump, suppressor, eluent
generator, and CR-TC. The flow rate, suppressor
current, and eluent concentration settings that were in
effect when the system was shut down are restored.
1. Verify that the pump is on and set to the correct flow rate.
2. Verify that the suppressor current is on and that the setting is correct.
3. Verify that the eluent generator is turned on and the correct concentration is
set. Verify that the CR-TC is turned on.
4. Verify that the cell heater is set to the correct value.