Oritron ICS-2100 Home Theater System User Manual


 
2 Features
Doc. 065291-01 3/09 33
Select the Options tab (see Figure 2-14).
.
Figure 2-14. Server Configuration Properties: Degas Options
Degas Options
Always Off: The degas pump is always off.
Always On: The degas pump is always on. This setting is for test
purposes by a Dionex Service Representative. Do not use this setting
for routine operation.
Cycle: The degas pump cycles on and off according to the times
specified in the
Cycle On and Off fields. Cycle On specifies for how
long the degas pump runs during a cycle.
Cycle Off specifies the time
between cycles.
Monitor: (default mode) The ICS-2100 monitors the degas pressure
reading and turns the degas pump on and off as required.