5•Service
Doc. 065291-01 3/09 151
11. Lift the EluGen
cartridge out of the
holder, turn it over,
and set it in the
service area.
12. Disconnect the
backpressure coil
from the EluGen
cartridge
OUTLET
port
.
13. Connect the EGC
OUT
line from the
holder to the
OUTLET port (see
Figure 5-45
).
Figure 5-45. EluGen Cartridge Inlet and Outlet Connections
14. Turn over the EluGen cartridge (fittings facing down). Check for
bubbles. If necessary, shake and tap the EluGen cartridge to remove
bubbles.
15. Reinstall the EluGen cartridge in the holder.
16. Reconnect the EluGen cartridge cable.
17. Reconnect the CR-TC cable.
18. If you disabled the
Automatically turn on with pump option in
Step 3
, select the option.
5.19.2 Replacing a K
2
CO
3
Cartridge
1. Turn off the pump from the ICS-2100 Control panel in Chromeleon
or from the ICS-2100 touch screen
HOME page.
2. Disconnect the EluGen cartridge cable from the
EGC-1 connector on
the ICS-2100 top cover.
3. If an EPM is installed, disconnect the EPM cable from the
EGC-2
connector.
EGC OUT
OUTLET
Port
EGC IN
INLET
Port