E.12 How do I know when a column is dirty?
See the troubleshooting section of the column manual.
E.13 How do I clean a column?
See the troubleshooting section of the column manual.
E.14 Why is the conductivity high?
Possible reasons for high conductivity include:
The suppressor is not on. Turn on the suppressor from the Chromeleon
Control panel or the ICS-2100 touch screen
SUPPRESSOR page.
The suppressor needs regeneration. See the suppressor manual for
troubleshooting information. The suppressor manual is included on the
Dionex Reference Library CD-ROM (P/N 053891).
The cell is out of calibration. See Section 5.1.3 for calibration instructions.
See Section 4.13 for additional troubleshooting information.
E.15 How do I configure and operate the auxiliary valve?
See Installing the ICS-1100/ICS-1600/ICS-2100 Auxiliary Valve (Document No.
065288), provided in the Auxiliary Valve Kit.