1. Locate the yellow 0.5 mL/min, 7 MPa (1000 psi) backpressure coil
(P/N 053765) and 10-32 to 1/4-28 coupler (P/N 042806) in the ICS-
2100 Ship Kit (P/N 064375).
2. Disconnect the
TO INJ VALVE IN - P/ELUENT OUT line from the EGC
CO
3
Mixer inlet.
3. Connect one end of the backpressure coil to the
TO INJ VALVE IN -
P/ELUENT OUT
line.
4. Connect the 10-32 to 1/4-28 coupler to the free end of the
backpressure coil.
5. Disconnect the
TO SRS/AES REGEN OUT line from the suppressor
REGEN OUT port.
6. Connect the
TO SRS/AES REGEN OUT line to the 10-32 to 1/4-28
coupler (see Figure 5-58
).
NOTE The backpressure coil connection in Figure 5-58
is temporary; it is required only during the
conditioning procedure described in the next
section.
Figure 5-58. Backpressure Coil Connections for Conditioning the EPM
with Carbonate/Bicarbonate Eluent
10-32 to
1/4-28
coupler
TO INJ
VALVE
IN - P/
ELUENT OUT
Backpressure
Coil
TO SRS/AES
REGEN OUT
10-32 to
10-32 coupler