ICS-2100 Ion Chromatography System
212 Doc. 065291-01 3/09
B.4 Home Page
B.4.1 Home Page Pump Controls
Pressure Display
The HOME page indicates the current pressure reading from
the pressure transducer. The pressure unit can be set to psi
(the factory default), MPa, or bar (see Section B.7.1
).
The minimum and maximum pressure limits are displayed
below the current pressure reading. If the pressure reading
goes outside these limits, the pump is stopped and an error
message appears in the Chromeleon Audit Trail and on the
touch screen. See Section 4.2
for troubleshooting
information. See Section B.7.1
for how to set the pressure
limits.
Figure B-7. ICS-2100 Touch Screen Home Page
Edit Field
Command
Button
(selected)
Command
Button
(unselected)
Status Field