details, refer to the EluGen cartridge manual. Cartridge manuals are
included on the Dionex Reference Library CD-ROM (P/N 053891).
EluGen Cartridge Holder
The EluGen cartridge is installed in a cartridge holder, which mounts on
the ICS-2100 top cover (see Figure 2-7
). The cartridge holder also houses
the high-pressure degas tubing assembly.
The ICS-2100 can control up to two EluGen cartridges. However, when
operating with an EGC II K
2
CO
3
EluGen Cartridge and the pH Modifier,
only one cartridge can be installed, because the pH Modifier must be
connected to the other EGC power supply (
EGC-1 or EGC-2).
EluGen
Cartridges
Eluent Concentration Range Comment
K
2
CO
3
/EPM
Electrolytic
pH Modifier
0.1 to 15 mM at 0.1 to 1.0 mL/min
flow
0.1 to X mM at 1.0 to 2.0 mL/min
flow
X = 15/flow
The total of the eluent
concentrations from
both the EluGen
cartridge and the EPM
must not exceed the
specified range. The
EPM concentration
must not exceed
10 mM. See
Generating
Carbonate/Bicarbonate
Eluent” on page 41 for
details.
KOH/KOH
KOH/MSA
KOH/NaOH
MSA/MSA
MSA/NaOH
NaOH/NaOH
0.1 to 50 mM at 0.1 to 1.0 mL/min
flow
0.1 to X mM at 1.0 to 3.0 mL/min
flow
X = 50/flow
The eluent
concentration range for
each cartridge is cut by
50%.
LiOH/LiOH 0.1 to 40 mM at 0.1 to 1.0 mL/min
flow
0.1 to X mM at 1.0 to 3.0 mL/min
flow
X = 40/flow
The eluent
concentration range for
each cartridge is cut by
50%.
Table 2-3. Eluent Concentration Ranges for Linked
Dual-Cartridge Configurations