9. Enter the column heater calibration values in Chromeleon or on the ICS-2100
touch screen
DIAGNOSTIC page as follows.
To enter the calibration values in Chromeleon:
a. Open the Wellness Panel (see Section 5.1.1
) and click Calibration
Details.
b. Under Column Heater Calibration Values, enter the Column Heater
Offset and Column Heater Slope values recorded in Step 4
.
c. In the field next to the Download button, verify that Current is selected
and then click Download to download the new calibration values to the
ICS-2100.
d. To ensure the new calibration values are recorded in the ICS-2100
memory, do not turn off the ICS-2100 power for at least 1 minute after
downloading the values.
To enter the calibration values on the ICS-2100 touch screen
DIAGNOSTIC
page:
a. Go to the
DIAGNOSTIC page and touch Column Heater Cal Values.
b. On the
COLUMN HEATER CAL VALUES page (see Figure 5-24), enter the
Calibrated Offset and Calibrated Slope values recorded in Step 4.
Figure 5-24. Column Heater Cal Values Page