Doc. 065291-01 3/09 143
3. Replace the two fuses in the holder with new IEC 127 fast-blow fuses rated
3.15 amps (P/N 954745). Dionex recommends always replacing both fuses.
4. Reinsert the fuse holder into its compartment and push in until the tab clicks
in place.
5. Reconnect the main power cord and turn on the power.
5.19 Replacing an EluGen Cartridge
The procedure for replacing an EluGen cartridge differs, depending on the type of
cartridge:
To replace a KOH, LiOH, MSA, or NaOH EluGen cartridge, go to
Section 5.19.1
.
To replace a K
2
CO
3
cartridge, go to Section 5.19.2.
5.19.1 Replacing a KOH, LiOH, MSA, or NaOH EluGen Cartridge
1. Turn off the pump from the ICS-2100 Control panel in Chromeleon
or from the ICS-2100 touch screen
HOME page.
2. Disconnect the EluGen cartridge cable from the
EGC-1 connector on
the ICS-2100 top cover.
3. Disconnect the CR-TC cable from the
CR-TC connector.
The EluGen cartridge contains one of the following: a corrosive base
(KOH, LiOH, or NaOH), or a corrosive acid (MSA). Wear protective
eyewear and gloves when handling the cartridge.
La cartouche d'EluGen contient un de ce qui suit: une base corrosive
(KOH, LiOH, ou NaOH), ou un acide corrosif (MSA). Porter des
lunettes et des gants protectives en manipulant la cartouche.
Die EluGen-Kartusche enthält eine korrosive Base (KOH, LiOH oder
NaOH), oder eine korrosive Säure (MSA). Tragen Sie daher beim
Umgang mit der Kartusche eine Schutzbrille und Handschuhe.