Status alarms and Status Alarm Severity (continued)Table 5-2:
Alarm code Status message Default severity Notes Configurable?
A103 Data Loss Possible (Totals
and Inventories)
Informational Applies only to flowmeters with the
standard core processor.
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A104 Calibration in Progress Informational Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A105 Slug Flow Informational Yes
A106 Burst Mode Enabled Informational Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A107 Power Reset Occurred Informational Normal transmitter behavior; oc-
curs after every power cycle.
Yes
A108 Basic Event 1 On Informational Applies only to basic events. Yes
A109 Basic Event 2 On Informational Applies only to basic events. Yes
A110 Frequency Output Satura-
ted
Informational Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A111 Frequency Output Fixed Informational Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A112 Upgrade Transmitter Soft-
ware
Informational Applies only to systems with trans-
mitter software earlier than v5.0.
Yes
A113 mA Output 2 Saturated Informational Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A114 mA Output 2 Fixed Informational Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A115 No External Input or Polled
Data
Informational Yes
A118 Discrete Output 1 Fixed Informational Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A119 Discrete Output 2 Fixed Informational Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A131 Meter Verification in Pro-
gress: Outputs to Last
Measured Value
Informational Applies only to transmitters with
Smart Meter Verification.
Yes
A132 Sensor Simulation Active Informational Applies only to flowmeters with the
enhanced core processor.
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A141 DDC trigger(s) have com-
pleted
Informational Applies only to flowmeters with the
enhanced core processor.
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
Configure device options and preferences
Configuration and Use Manual 75