Smart Meter Verification – Top-level menuFigure 9-3:
Run Verify Read Results Schedule Verify Exit
Down Down Down
Down Select
Select Select Select
*This option is displayed only if the transmitter is connected
to an enhanced core processor (V3.6 or higher) and the
meter verification software is installed on the transmitter.
Online Verify*
Select
Down
Select
Process variable
display
Up
Select
2. Choose Run Verify.
3. Choose the desired output behavior.
Option Description
Continue Meas-
ure
During the test, all outputs will continue to report their assigned process
variable. The test will run for approximately 90 seconds.
Fault Value During the test, all outputs will report the last measured value of their as-
signed process variable. The test will run for approximately 140 seconds.
Last Value During the test, all outputs will go to their configured fault action. The test
will run for approximately 140 seconds.
While the test is in progress, dots traverse the display and test progress is shown.
Postrequisites
View the test results and take any appropriate actions.
Measurement support
138 Micro Motion
®