Option Description
Fault Actions when fault is detected:
• The alarm is posted to the Alert List.
• Outputs go to the configured fault action (after Fault Timeout has expired, if
applicable).
• Digital communications go to the configured fault action (after Fault Timeout
has expired, if applicable).
• The status LED (if available) changes to red or yellow (depending on alarm se-
verity).
Actions when alarm clears:
• Outputs return to normal behavior.
• Digital communications return to normal behavior.
• The status LED (if available) returns to green and may or may not flash.
Informa-
tional
Actions when fault is detected:
• The alarm is posted to the Alert List.
• The status LED (if available) changes to red or yellow (depending on alarm se-
verity).
Actions when alarm clears:
• The status LED (if available) returns to green and may or may not flash.
Ignore No action
Status alarms and options for Status Alarm Severity
Status alarms and Status Alarm SeverityTable 5-2:
Alarm code Status message Default severity Notes Configurable?
A001 EEPROM Error (Core Pro-
cessor)
Fault No
A002 RAM Error (Core Processor) Fault No
A003 No Sensor Response Fault Yes
A004 Temperature Overrange Fault No
A005 Mass Flow Rate Overrange Fault Yes
A006 Characterization Required Fault Yes
A008 Density Overrange Fault Yes
A009 Transmitter Initializing/
Warming Up
Fault Yes
A010 Calibration Failure Fault No
A011 Zero Calibration Failed:
Low
Fault Yes
A012 Zero Calibration Failed:
High
Fault Yes
A013 Zero Calibration Failed:
Unstable
Fault Yes
Configure device options and preferences