Display Variable 1 will automatically be set to match mA Output Process Variable for the primary
mA output. If you change the configuration of mA Output Process Variable, Display Variable 1
will be updated automatically.
5.1.3 Configure the precision of variables shown on the
display
Display (standard) Not available
Chinese-language
display
Offline Maintain > Configuration > Display > Num of Decimals
ProLink II ProLink > Configuration > Display > Display Precision
ProLink III Device Tools > Configuration > Transmitter Display > Display Variables
Field Communicator Configure > Manual Setup > Display > Decimal Places
Overview
Setting Display Precision determines the precision (number of decimal places) shown on the
display. You can set Display Precision independently for each variable.
Setting Display Precision does not affect the actual value of the process variable.
Procedure
1. Select a process variable.
2. Set Display Precision to the number of decimal places you want shown when the
process variable appears on the display.
For temperature and density process variables, the default value is 2 decimal places.
For all other process variables, the default value is 4 decimal places. The range is
0 to 5.
Tip
The lower the selected precision, the greater the process change must be for it to be reflected
on the display. Do not set Display Precision value too low or too high to be useful.
5.1.4 Configure the refresh rate of data shown on the display
Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > RATE
Chinese-language
display
Offline Maintain > Configuration > Display > Display Rate
ProLink II ProLink > Configuration > Display > Update Period
ProLink III Device Tools > Configuration > Transmitter Display > Display Variables
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Refresh Rate
Configure device options and preferences