Smart Meter Verification – Top-level menuFigure 9-11:
Run Verify Read Results Schedule Verify Exit
Down Down Down
Down Select
Select Select Select
*This option is displayed only if the transmitter is connected
to an enhanced core processor (V3.6 or higher) and the
meter verification software is installed on the transmitter.
Online Verify*
Select
Down
Select
Process variable
display
Up
Select
2. Scroll to Schedule Verify and press Select.
3. To schedule a single test or the first test in recurring execution:
a. Scroll to Set Next and press Select.
b. Enter the number of hours that the transmitter will wait before beginning the
test.
4. To schedule recurring execution:
a. Scroll to Set Recur and press Select.
b. Enter the number of hours that will elapse between tests.
5. To disable scheduled execution:
• To disable execution of a single scheduled test, set Set Next to 0.
• To disable recurring execution, set Set Recur to 0.
• To disable all scheduled execution, choose Turn Off Schedule when you enter the
Smart Meter Verification menu.
Measurement support
150 Micro Motion
®
Model 1700 Transmitters with Analog Outputs