Smart Meter Verification flowchart: Scheduling test execution using the Chinese-
language display
Scheduling Smart Meter Verification test execution using the Chinese-language displayFigure 9-12:
Schedule Verify
Select
Save?/Yes
Turn Off Schedule?/Yes
Set Next
Select
Select
Schedule set?
Yes
Schedule deleted
Up
Hours Left
SelectDown
Hours Left xx
Set Recurrence
No
Schedule is Off
xx HOURS
Select
Up
No Yes
Select
xx HOURS
Select
Save?/Yes
Up
No Yes
Down
Down
Exit
Up Select
Select
Down
Manage scheduled test execution using ProLink II
1. Choose Tools > Meter Verification > Schedule Meter Verification.
2. To schedule a single test or the first test in recurring execution, specify a value for
Hours Until Next Run.
3. To schedule recurring execution, specify a value for Hours Between Recurring Runs.
4. To disable scheduled execution:
To disable execution of a single scheduled test, set Hours Until Next Run to 0.
To disable recurring execution, set Hours Between Recurring Runs to 0.
To disable all scheduled execution, click Turn Off Schedule.
Manage scheduled test execution using ProLink III
1. Choose Device Tools > Diagnostics > Meter Verification > Schedule Meter Verification.
2. To schedule a single test or the first test in recurring execution, specify a value for
Hours Until Next Run.
Measurement support
Configuration and Use Manual 151