Emerson Model 1700 Satellite Radio User Manual


 
Recent Alerts
Alarm data in transmitter memoryTable 8-2:
Alarm data structure
Transmitter action if condition occurs
Contents Clearing
Alert List As determined by the alarm status bits, a list
of:
All currently active alarms
All previously active alarms that have not
been acknowledged
Cleared and regenerated with every transmit-
ter power cycle
Alert Statistics One record for each alarm (by alarm number)
that has occurred since the last master reset.
Each record contains:
A count of the number of occurrences
Timestamps for the most recent posting
and clearing
Not cleared; maintained across transmitter
power cycles
Recent Alerts 50 most recent alarm postings or alarm clear-
ings
Not cleared; maintained across transmitter
power cycles
8.5 Read totalizer and inventory values
Display (standard) To read a totalizer or inventory value from the display, it must be configured as a display
variable.
Chinese-language
display
To read a totalizer or inventory value from the Chinese-language display, it must be con-
figured as a display variable.
ProLink II ProLink > Totalizer Control
ProLink III View the desired variable on the main screen under Process Variables.
Field Communicator Service Tools > Variables > Totalizer Control
Overview
Totalizers keep track of the total amount of mass or volume measured by the transmitter
since the last totalizer reset. Inventories keep track of the total amount of mass or volume
measured by the transmitter since the last inventory reset.
Tip
You can use the inventories to keep a running total of mass or volume across multiple totalizer
resets.
Transmitter operation
128 Micro Motion
®
Model 1700 Transmitters with Analog Outputs