Prerequisites
The temperature calibration is a two-part procedure: temperature offset calibration and
temperature slope calibration. The two parts must be performed without interruption, in
the order shown. Ensure that you are prepared to complete the process without
interruption.
Important
Consult Micro Motion before performing a temperature calibration. Under normal circumstances,
the temperature circuit is stable and should not need an adjustment.
Procedure
See #unique_248/PL3TempCalFig.
Temperature calibration using ProLink IIIFigure 9-20:
Enter temperature of low-
temperature fluid
Temperature Offset calibration
Wait until sensor achieves
thermal equilibrium
Fill sensor with low-
temperature fluid
Enter temperature of high-
temperature fluid
Temperature Slope calibration
Start Calibration
Wait until sensor achieves
thermal equilibrium
Fill sensor with high-
temperature fluid
Done
Device Tools >
Calibration >
Temperature Calibration >
Temperature Calibration - Offset
Start Calibration
Device Tools >
Calibration >
Temperature Calibration >
Temperature Calibration - Slope