Status alarms and Status Alarm Severity (continued)Table 5-2:
Alarm code Status message Default severity Notes Configurable?
A014 Transmitter Failure Fault No
A016 Sensor RTD Failure Fault Yes
A017 T-Series RTD Failure Fault Yes
A018 EEPROM Error (Transmit-
ter)
Fault No
A019 RAM Error (Transmitter) Fault No
A020 No Flow Cal Value Fault Yes
A021 Incorrect Sensor Type (K1) Fault No
A022 Configuration Database
Corrupt (Core Processor)
Fault Applies only to flowmeters with the
standard core processor.
No
A023 Internal Totals Corrupt
(Core Processor)
Fault Applies only to flowmeters with the
standard core processor.
No
A024 Program Corrupt (Core
Processor)
Fault Applies only to flowmeters with the
standard core processor.
No
A025 Boot Sector Fault (Core
Processor)
Fault Applies only to flowmeters with the
standard core processor.
No
A026 Sensor/Transmitter Com-
munications Failure
Fault No
A028 Core Processor Write Fail-
ure
Fault No
A031 Low Power Fault Applies only to flowmeters with the
enhanced core processor.
No
A032 Meter Verification in Pro-
gress: Outputs to Fault
Varies Applies only to transmitters with
Smart Meter Verification.
If outputs are set to Last Measured
Value, severity is Info. If outputs are
set to Fault, severity is Fault.
No
A033 Insufficient Right/Left Pick-
off Signal
Fault Applies only to flowmeters with the
enhanced core processor.
Yes
A034 Meter Verification Failed Informational Applies only to transmitters with
Smart Meter Verification.
Yes
A035 Meter Verification Aborted Informational Applies only to transmitters with
Smart Meter Verification.
Yes
A100 mA Output 1 Saturated Informational Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A101 mA Output 1 Fixed Informational Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A102 Drive Overrange Informational Yes
Configure device options and preferences
74 Micro Motion
®
Model 1700 Transmitters with Analog Outputs