1. Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and
checks for error conditions. During the power-up sequence, Alarm A009 is active.
This alarm should clear automatically when the power-up sequence is complete.
2. Check the status LED on the transmitter.
Transmitter status reported by status LEDTable 2-1:
LED state Description Recommendation
Green No alarms are active. Continue with configuration or process meas-
urement.
Flashing green
(1)
No alarms are active. One or more previously
active alarms have not been acknowledged.
Continue with configuration or process meas-
urement. If you choose, you can acknowledge
the alarms.
Yellow One or more low-severity alarms are active,
and have been acknowledged.
A low-severity alarm condition does not affect
measurement accuracy or output behavior.
You can continue with configuration or proc-
ess measurement. If you choose, you can iden-
tify and resolve the alarm condition.
Flashing yellow
(2)
One or more low-severity alarms are active
and have not been acknowledged.
A low-severity alarm condition does not affect
measurement accuracy or output behavior.
You can continue with configuration or proc-
ess measurement. If you choose, you can iden-
tify and resolve the alarm condition. You may
also acknowledge the alarm.
Red One or more high-severity alarms are active,
and have been acknowledged.
A high-severity alarm condition affects meas-
urement accuracy and output behavior. Re-
solve the alarm condition before continuing.
Flashing red
(3)
One or more high-severity alarms are active
and have not been acknowledged.
A high-severity alarm condition affects meas-
urement accuracy and output behavior. Re-
solve the alarm condition before continuing.
You may also acknowledge the alarm.
Postrequisites
For information on viewing the list of active alarms, see Section 8.4.
For information on individual alarms and suggested resolutions, see Section 10.2.
(1) If Status LED Blinking is disabled, the LED will show solid green rather than flashing.
(2) If Status LED Blinking is disabled, the LED will show solid yellow rather than flashing.
(3) If Status LED Blinking is disabled, the LED will show solid red rather than flashing.
Quick start
6 Micro Motion
®
Model 1700 Transmitters with Analog Outputs