D3 or D3 and D4 density calibration using the Field CommunicatorFigure 9-18:
Enter density of D3 fluid
Density Calibration
Complete message
Calibration in Progress
message
D3 calibration
Close shutoff valve
downstream from sensor
Fill sensor with D3 fluid Fill sensor with D4 fluid
D4 calibration
OK
Dens Pt 3 T-Series
Calibration method
executes
OK
Home
Enter density of D4 fluid
Density Calibration
Complete message
Calibration in Progress
message
OK
Dens Pt 4 T-Series
Calibration method
executes
OK
Home
Done
On-Line Menu >
Service Tools >
Maintenance >
Density Calibration
Service Tools >
Maintenance >
Density Calibration
Done
9.7 Perform temperature calibration
Temperature calibration establishes the relationship between the temperature of the
calibration fluids and the signal produced by the sensor.
9.7.1 Perform temperature calibration using ProLink II
Temperature calibration establishes the relationship between the temperature of the
calibration fluids and the signal produced by the sensor.
Measurement support
170 Micro Motion
®
Model 1700 Transmitters with Analog Outputs