Emerson Process Management 1500 Satellite Radio User Manual


 
90 Micro Motion
®
Model 1500 Transmitters with the Filling and Dosing Application
Measurement Performance
10.6 Performing temperature calibration
Temperature calibration is a two-part procedure: temperature offset calibration and temperature slope
calibration. The entire procedure must be completed without interruption.
You can calibrate for temperature with ProLink II. See Figure 10-4.
Figure 10-4 Temperature calibration – ProLink II
Enter temperature of low-
temperature fluid
Temperature Offset calibration
Do Cal
Wait until sensor achieves
thermal equilibrium
Fill sensor with low-
temperature fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
Close
Enter temperature of high-
temperature fluid
Temperature Slope calibration
Do Cal
Wait until sensor achieves
thermal equilibrium
Fill sensor with high-
temperature fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
Close
Done
ProLink Menu >
Calibration >
Temp offset cal
ProLink Menu >
Calibration >
Temp slope cal