Model 1500 Transmitters with the Filling and Dosing Application
Optional Transmitter Configuration
For a list of all status alarms and default severity levels, see Table 6-5. (For more information on
status alarms, including possible causes and troubleshooting suggestions, see Section 11.10.)
Table 6-5 Status alarms and severity levels
Alarm code ProLink II message
Default
severity Configurable?
Affected by
fault timeout?
A001 CP EEPROM Failure Fault No No
A002 CP RAM Failure Fault No No
A003 Sensor Failure Fault No Yes
A004 Temp Out of Range Fault No Yes
A005 Mass Flow Overrange Fault No Yes
A006 Characterize Meter Fault No No
A008 Density Out of Range Fault No Yes
A009 Xmtr Initializing Fault No No
A010 Calibration Failure Fault No No
A011 Cal Fail, Too Low Fault No No
A012 Cal Fail, Too High Fault No No
A013 Cal Fail, Too Noisy Fault No No
A014 Transmitter Error Fault No No
A016 Sensor RTD Error Fault No Yes
A017 Meter RTD Error Fault No Yes
A018 EEPROM Failure Fault No No
A019 RAM Failure Fault No No
A020 Cal Factors Missing Fault No No
A021 Sensor Type Incorrect Fault No No
A022
(1)
CP Configuration Failure Fault No No
A023
(1)
CP Totals Failure Fault No No
A024
(1)
CP Program Corrupt Fault No No
A025
(1)
CP Boot Program Fault Fault No No
A026 Xmtr Comm Problem Fault No No
A028 Comm Problem Fault No No
A032
(2)
Meter Verification/Outputs In Fault Fault No No
A100 mA 1 Saturated Info Yes No
A101 mA 1 Fixed Info Yes No
A102 Drive Overrange/Partially Full Tube Info Yes No
A103
(1)
Data Loss Possible Info Yes No
A104 Cal in Progress Info Yes No
A105 Slug Flow Info Yes No
A107 Power Reset Info Yes No
A108 Event 1 On Info Yes No
A109 Event 2 On Info Yes No
A112 Upgrade Software Info Yes No
A115 External Input Error Info Yes No