Emerson Process Management 1500 Satellite Radio User Manual


 
Configuration and Use Manual 89
Measurement Performance
Measurement Performance DefaultsTroubleshootingCompensation
Figure 10-2 D1 and D2 density calibration – ProLink II
Figure 10-3 D3 or D3 and D4 density calibration – ProLink II
Enter density of D1 fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
D1 calibration
Close shutoff valve
downstream from sensor
Fill sensor with D1 fluid Fill sensor with D2 fluid
Close
Enter density of D2 fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
D2 calibration
Close
Done
Do Cal Do Cal
ProLink Menu >
Calibration >
Density cal – Point 1
ProLink Menu >
Calibration >
Density cal – Point 2
Enter density of D3 fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
D3 calibration
Close shutoff valve
downstream from sensor
Fill sensor with D3 fluid
Close
Enter density of D4 fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
D4 calibration
Close
Done
Do Cal Do Cal
Done
ProLink Menu >
Calibration >
Density cal – Point 3
Fill sensor with D4 fluid
ProLink Menu >
Calibration >
Density cal – Point 4