Emerson Process Management 1500 Satellite Radio User Manual


 
Configuration and Use Manual 83
Measurement Performance
Measurement Performance DefaultsTroubleshootingCompensation
10.2.4 Comparison and recommendations
When choosing among meter verification, meter validation, and calibration, consider the following
factors:
Process interruption
- Meter verification requires approximately four minutes to perform. During these four
minutes, flow can continue (provided sufficient stability is maintained); however, outputs
will not report process data.
- Meter validation for density does not interrupt the process at all. However, meter
validation for mass flow or volume flow requires process down-time for the length of the
test.
- Calibration requires process down-time. In addition, density and temperature calibration
require replacing the process fluid with low-density and high density fluids, or
low-temperature and high-temperature fluids.
External measurement requirements
- Meter verification does not require external measurements.
- Zero calibration does not require external measurements.
- Density calibration, temperature calibration, and meter validation require external
measurements. For good results, the external measurement must be highly accurate.
Measurement adjustment
- Meter verification is an indicator of sensor condition, but does not change flowmeter
internal measurement in any way.
- Meter validation does not change flowmeter internal measurement in any way. If you
decide to adjust a meter factor as a result of a meter validation procedure, only the reported
measurement is changed – the base measurement is not changed. You can always reverse
the change by returning the meter factor to its previous value.
- Calibration changes the transmitter’s interpretation of process data, and accordingly
changes the base measurement. If you perform a zero calibration, you can restore the
factory zero at a later time. You cannot return to the previous zero (if different from the
factory zero), density calibration values, or temperature calibration values unless you have
manually recorded them.
Micro Motion recommends obtaining the meter verification transmitter option and performing meter
verification on a regular basis.
10.3 Performing meter verification
Note: To use meter verification, the transmitter must be paired with an enhanced core processor, and
the meter verification option must be purchased for the transmitter.
The meter verification procedure can be performed on any process fluid. It is not necessary to match
factory conditions. Meter verification is not affected by any parameters configured for flow, density,
or temperature.
During the test, process conditions must be stable. To maximize stability:
Maintain a constant temperature and pressure.
Avoid changes to fluid composition (e.g., two-phase flow, settling, etc.).
Maintain a constant flow. For higher test certainty, reduce or stop flow.