Model 1500 Transmitters with the Filling and Dosing Application
Required Transmitter Configuration
View the trend chart for these initial tests. By default, the specification uncertainty limit is set at
±4.0%, which will avoid false Fail/Caution results over the entire range of specified process
conditions. If you observe a structural integrity variation greater than 4% due to normal process
conditions, you may adjust the specification uncertainty limit to match your process variation. To
avoid false Fail/Caution results, it is advisable to set the specification uncertainty limit to
approximately twice the variation due to the effect of normal process conditions.
In order to perform this baseline analysis, you will need the enhanced meter verification capabilities
of ProLink II v2.5 or later. Refer to the manual entitled ProLink
®
II Software for Micro Motion
®
Transmitters: Installation and Use, P/N 20001909, Rev D or later.