Digitizers Command Reference 85Chapter 3
[SENSe:]
The SENSe command subsystem is used to change low-level parameters
such as voltage range, sweep and sweep offset points and to set the
reference oscillator source and frequency. It is also used to obtain
measurement data from the module.
Subsystem Syntax [SENSe:]
DATA? <rdgs_per_channel>[,channel_list]
DATA:ALL? <rdgs_per_channel>
DATA:COUNt?
DATA:CVTable? [channel_list]
ROSCillator:EXTernal:FREQuency <freq>l
ROSCillator:EXTernal:FREQuency?
ROSCillator:SOURce INTernal | EXTernal
ROSCillator:SOURce?
SWEep:OFFSet:POINts <neg_value> | MIN | MAX
SWEep:OFFSet:POINts? MIN | MAX
SWEep:POINts <neg_value> | MIN | MAX
SWEep:POINts? MIN | MAX
VOLTage[<channel>][:DC]:RANGe <range> | MIN | MAX
VOLTage[<channel>][:DC]:RANGe?
VOLTage[<channel>][:DC]:RESolution?
[SENSe:]DATA?
[SENSe:]DATA? <rdgs_per_channel>[,channel_list] returns voltage formatted
data from all channels (default) or only from the specified channel list.
<channel_list> has the form (@1) or (@2), (@1,2), (@1:4) or (@1,2,3,4).
For specific channels, but not all, the format is (@1,3,4).
Parameters
Name Type Range of Values Default Value
rdgs_per_
channel
numeric 1 to MAX samples
depends on size of RAM
on module
(see SAMPle:COUNt)
none
channel_list numeric 1-2 (E1563A)
1-4 (E1564A)
N/A